Assessment of the Eppendorf ERIS analyser
1985

Assessment of the Eppendorf ERIS Analyser

Sample size: 100 publication Evidence: moderate

Author Information

Author(s): S. J. Callaghan, R. Singer, J. M. White, C. G. Fraser

Primary Institution: Department of Biochemical Medicine, Ninewells Hospital and Medical School

Hypothesis

The Eppendorf ERIS analyser is suitable as a partial replacement for existing fast analysers at Ninewells Hospital.

Conclusion

The Eppendorf ERIS proved to be a reliable instrument with acceptable performance characteristics for the analytes studied.

Supporting Evidence

  • The Eppendorf ERIS can perform up to 20 tests on each sample.
  • It produces one result every 9 seconds after a 10-minute lag.
  • Calibration is recommended weekly or when changes occur.
  • Carry-over for most tests was negligible.
  • Imprecision was acceptable for clinical purposes.

Takeaway

The Eppendorf ERIS can run many tests quickly and accurately, making it great for hospitals.

Methodology

The performance characteristics of various assays were evaluated using the Eppendorf ERIS and compared to a parallel fast analyser.

Potential Biases

The Eppendorf ERIS showed some positive proportional bias in iron and GGT results.

Limitations

Extensive studies on linearity, interfering substances, and recovery were not carried out due to time constraints.

Participant Demographics

Patients' sera were selected to cover the analytical range and represent typical specimens.

Statistical Information

P-Value

0.001 > p > 0.01

Statistical Significance

p<0.05

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication