Evaluation of the Gemsaec 3E centrifugal analyser
1981
Evaluation of the Gemsaec 3E Centrifugal Analyser
Sample size: 100
publication
Evidence: moderate
Author Information
Author(s): Peter S. West, Jennifer A. Nisbet, John A. Owen
Primary Institution: Department of Chemical Pathology, St. George's Hospital, London SW17, UK.
Conclusion
The precision and accuracy of the Gemsaec 3E centrifugal analyser are adequate for routine clinical chemistry.
Supporting Evidence
- The Gemsaec 3E provides a generally acceptable level of analytical performance.
- Precision was measured at low, medium, and high assay levels with coefficients of variation less than 5% for most tests.
- Results from the Gemsaec were compared with routine laboratory methods showing good agreement.
Takeaway
The Gemsaec 3E is a machine that helps doctors test blood samples quickly and accurately, making it easier to check for diseases.
Methodology
The study involved evaluating the analytical performance of the Gemsaec 3E using various tests and comparing results with routine laboratory methods.
Limitations
The evaluation was based on only 8 different tests, which may not represent the performance across all tests.