An automatic collector for density gradients
1980

Automatic Collector for Density Gradients

publication Evidence: moderate

Author Information

Author(s): Rex Mottershead

Primary Institution: Medical Research Council Cell Mutation Unit, University of Sussex

Hypothesis

The automatic collector improves the collection of density gradients by eliminating manual counting and timing.

Conclusion

The automatic collector is as accurate as manual methods and significantly speeds up the process of collecting density gradients.

Supporting Evidence

  • The automatic collector eliminates the need for manual counting and timing, reducing operator fatigue.
  • The profiles of gradients collected manually and automatically were very similar, indicating comparable accuracy.
  • The machine has been used routinely for over a year without revealing fundamental problems.

Takeaway

This study shows a new machine that helps scientists collect liquid samples faster and without mistakes from tired hands.

Methodology

The collector uses a peristaltic pump to transfer density gradient contents onto paper tape, and its performance was compared to manual collection methods.

Limitations

Minor faults can occur due to blockages in the tubing, but no fundamental problems were reported.