Erythrocyte transketolase activity measured on the Gemstar discrete analyser
1984

Measuring Erythrocyte Transketolase Activity with Gemstar Analyser

Sample size: 22 publication Evidence: moderate

Author Information

Author(s): J. E. Buttery, B. R. Chamberlain, C. R. Milner

Primary Institution: The Queen Elizabeth Hospital

Conclusion

The Gemstar analyser provides comparable results to the manual method for measuring erythrocyte transketolase activity.

Supporting Evidence

  • The correlation for the TK assay between Gemstar and manual methods was r=0.98.
  • The mean TK activity measured by Gemstar was 0.84 U/g Hb compared to 0.82 U/g Hb by the manual method.
  • The reference range for erythrocyte TK activity on the Gemstar is 0.6 to 1.3 U/g Hb.

Takeaway

This study shows a new machine can measure a specific enzyme in blood samples just as well as the old method, but faster.

Methodology

The study adapted a kinetic assay for erythrocyte transketolase to the Gemstar analyser and compared it with a manual method.

Limitations

The day-to-day precision could not be performed due to specimen instability.