An assessment of the performance of the Chemispek J200 in a hospital laboratory
1982

Performance Assessment of the Chemispek J200 Autoanalyser

Sample size: 107 publication Evidence: moderate

Author Information

Author(s): Gordon Dale, Andrew McGill, David R. Walton

Primary Institution: Department of Clinical Biochemistry, Newcastle General Hospital

Hypothesis

The Chemispek J200 will perform effectively in a hospital laboratory setting.

Conclusion

The Chemispek J200 has been shown to perform very well, being fast, precise, and accurate despite some initial mechanical issues.

Supporting Evidence

  • The Chemispek J200 can analyze up to 16 samples simultaneously.
  • It achieved a throughput of 120 specimens per hour.
  • Standardization showed high linearity for most channels.
  • Drift correction improved test results significantly.
  • Carry-over effects were minimized with correction factors.

Takeaway

The Chemispek J200 is a machine that helps test samples quickly and accurately, but it had some problems at first that have mostly been fixed.

Methodology

The study involved analyzing various serum samples using the Chemispek J200 and comparing results with established standards.

Potential Biases

Potential for erroneous results due to unflagged abnormal peaks in some channels.

Limitations

Mechanical issues with the gear-box and some channels experienced blockages, affecting performance.