Evaluation of the new 'C' parallel analyser
1980

Evaluation of the C Parallel Analyser

Sample size: 100 publication Evidence: moderate

Author Information

Author(s): Raija Puukka, M. Puukka

Primary Institution: Department of Clinical Chemistry, Oulu University Central Hospital

Hypothesis

The C Parallel Analyser is a new clinical analyser that can perform photometric equilibrium and kinetic analyses effectively.

Conclusion

The C Parallel Analyser has demonstrated good reliability and precision in clinical assays.

Supporting Evidence

  • The analyser can process 24 samples in 10 minutes for albumin and 25 minutes for other methods.
  • Correlation coefficients for the analyser's results compared to routine methods were over 0.99 for most tests.
  • Within-day precision was generally less than 3.0%.

Takeaway

The C Parallel Analyser is a machine that helps doctors quickly and accurately test blood samples for different substances.

Methodology

The analyser was evaluated by comparing its results with those from routine methods for various serum tests.

Limitations

The analyser has limitations in handling very rapid kinetic reactions and lacks a graphic printer.

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication