Evaluation of the new 'C' parallel analyser
1980
Evaluation of the C Parallel Analyser
Sample size: 100
publication
Evidence: moderate
Author Information
Author(s): Raija Puukka, M. Puukka
Primary Institution: Department of Clinical Chemistry, Oulu University Central Hospital
Hypothesis
The C Parallel Analyser is a new clinical analyser that can perform photometric equilibrium and kinetic analyses effectively.
Conclusion
The C Parallel Analyser has demonstrated good reliability and precision in clinical assays.
Supporting Evidence
- The analyser can process 24 samples in 10 minutes for albumin and 25 minutes for other methods.
- Correlation coefficients for the analyser's results compared to routine methods were over 0.99 for most tests.
- Within-day precision was generally less than 3.0%.
Takeaway
The C Parallel Analyser is a machine that helps doctors quickly and accurately test blood samples for different substances.
Methodology
The analyser was evaluated by comparing its results with those from routine methods for various serum tests.
Limitations
The analyser has limitations in handling very rapid kinetic reactions and lacks a graphic printer.