Evaluation of the Beckman Astra 8 analyser

Evaluation of the Beckman Astra 8 Analyser

publication

Author Information

Author(s): Lloyd et al.

Conclusion

The Astra 8 analyser is precise and accurate for measuring electrolytes, urea, glucose, and creatinine in serum and urine.

Supporting Evidence

  • The Astra system has overcome many problems associated with the kinetic Jaffe procedure.
  • It is easy to use and requires simple routine maintenance.
  • The RAPid kit allows for quick replacement of defective modules.
  • It can analyze 200-300 specimens per day.

Takeaway

The Astra 8 is a machine that helps test blood and urine samples quickly and accurately.

Potential Biases

The results may be affected by the deterioration of the reagent in the measuring cup.

Limitations

The analyser may give erroneously low results if urine samples are followed by serum samples, and this effect is not reproducible.

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication